The associated complaint devices were returned and evaluated.For part: 71674001: a visual inspection confirms that one device shows epoxy flaking inside profile and some dents and scratches.These devices are connecting with mating device without issue.At this time, we are unable to confirm stated failure of these devices.A clinical evaluation was completed for part number 7167001 & this compliant reports an intra-operative implant to implant connection malfunction.During the procedure the surgeon reported, ¿a conversion from the standard surgical procedure to a perfect circles technique to freehand lock the distal hole which prolonged the surgical procedure a little (the length of the delay is unknown)¿.According to the surgeon, ¿there was no patient injury from this¿.The associated implants were returned and evaluated.However, the device analysis did not indicate any mal-performance of the smith and nephew devices.Since there was no alleged harm to the patient, no further clinical/medical investigation warranted at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|