• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE Back to Search Results
Model Number 29525
Device Problem Cut In Material (2454)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).No specific issue was found in the service or complaint system for the return.Verified technician finding that the power cord was damaged and copper wiring was exposed.A pm/safety label on the front of the unit had the words "frayed cord" written over it in marker.Visual inspection of the cord found a cut in the outer insulation across the cord that had severed the internal black and white wires, exposing the inner copper.Power cord was removed and retained.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Upon triage, the technician found that the power cord was damaged and copper wiring was exposed.
 
Manufacturer Narrative
One kendall scd 700 sequential compression system was received with no complaint listed.The unit was triaged and service found that the power cord was damaged exposing copper wires.Visual inspection of the cord found a cut in the outer insulation across the cord that had severed the internal black and white wires, exposing the inner copper.The power cord was replaced.The potential root cause is customer misuse by running over the cord with hospital bed or due to the procedure used to unplug the unit and the procedure of wrapping of the cord around the bed hook.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Type of Device
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
Manufacturer (Section G)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN   201114
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7564694
MDR Text Key109889254
Report Number3006451981-2018-00413
Device Sequence Number1
Product Code JOW
UDI-Device Identifier10884521129566
UDI-Public10884521129566
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/12/2013
Device Model Number29525
Device Catalogue Number29525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2018
Initial Date FDA Received06/04/2018
Supplement Dates Manufacturer Received06/01/2018
Supplement Dates FDA Received06/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-