| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received at zoll medical corporation for evaluation.The reported problem was verified and attributed to a faulty transformer on the processor/bridge/pace board.The processor/bridge/pace board was replaced to remedy the problem.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self test for defib function.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Created in error et (b)(4) 2018.
|
| |
|
Event or Problem Description
|
|
Created in error et (b)(4) 2018.
|
| |
|