| Catalog Number |
5512-F-601 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348); Joint Dislocation (2374)
|
| Date of Event |
12/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been two other events for the reported lot.Both relate to the same patient.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
|
| |
|
Event or Problem Description
|
|
Patient's left knee dislocated and was treated with closed reduction under anesthesia (b)(6) 2017.Per exam notes: "patient states that he was in his recliner this morning and when he went to get up his knee popped, immediate pain, and can not bear weight.He also twisted his ankle last week.".Patient was due to go to the hospital to address the twisted ankle the afternoon of the day he dislocated.Operative report indicates dislocation was without fracture and that in performing the closed reduction, a "palpable pop" was noted with fluoroscopy revealing "acceptable reduction of the knee dislocation both in ap and lateral planes".
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: event executive summary an event regarding revision due to dislocation of a triathlon insert involving a triathlon ts femoral component was reported.The event was confirmed based on the clinician review of the medical records provided.Method & results: -device evaluation and results: not performed as product was not returned -medical records received and evaluation: clinician review of the medical records provided indicated "no examination of any explanted components except for the insert from the (b)(6)2018 revision surgery and no primary left total knee arthroplasty operative report or list of components are available for review.The medial position of the revision tibial component and the distal position of the joint line in the revision total knee arthroplasty likely resulted in increased stress to the ts insert stability resulting in two dislocations and a later anterior implant subluxation.There is no evidence that factors of implant design, manufacturing or materials were responsible for this clinical situation".-device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.-complaint history review: there have been three other similar events for the reported lot.Conclusion: clinician review of the medical records provided indicated "no examination of any explanted components except for the insert from the (b)(6)2018 revision surgery and no primary left total knee arthroplasty operative report or list of components are available for review.The medial position of the revision tibial component and the distal position of the joint line in the revision total knee arthroplasty likely resulted in increased stress to the ts insert stability resulting in two dislocations and a later anterior implant subluxation.There is no evidence that factors of implant design, manufacturing or materials were responsible for this clinical situation".No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Patient's left knee dislocated and was treated with closed reduction under anesthesia (b)(6)2017.Per exam notes: "patient states that he was in his recliner this morning and when he went to get up his knee popped, immediate pain, and can not bear weight.He also twisted his ankle last week.".Patient was due to go to the hospital to address the twisted ankle the afternoon of the day he dislocated.Operative report indicates dislocation was without fracture and that in performing the closed reduction, a "palpable pop" was noted with fluoroscopy revealing "acceptable reduction of the knee dislocation both in ap and lateral planes".Update: as per consulting clinician's medical review, "the medial position of the revision tibial component and the distal position of the joint line in the revision total knee arthroplasty likely resulted in increased stress to the ts insert stability resulting in two dislocations and a later anterior implant subluxation." the consulting clinician clarified that this conclusion reflects an alignment issue of the tibial baseplate and femoral component leading toward subsequent subluxation and dislocation.
|
| |
|