Model Number 15
Medical Device Problem Code Failure To Run On AC/DC (1001)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that when attempting to power on their device, it will not power on.The device will not power on when plugged into ac and the device has two fully charged batteries installed.There was no patient use associated with the reported event.
 
Event or Problem Description
The customer contacted physio-control to report that when attempting to power on their device, it will not power on.The device will not power on when plugged into ac and the device has two fully charged batteries installed.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control evaluated the customer's device but was unable to duplicate the reported power issue.Physio downloaded and reviewed the electronic records from the customer¿s device and verified the reported issue.Physio observed that the device was not communicating with the ac power adapter while in use in a king air 200 fixed wing air ambulance connected to that ambulances static inverter.Physio determined that the likely cause of the reported issue was due to a failure of the ac power adapter to communicate with the device as well as ac power line artifact from their "clipped sine wave" inverter on the airplane.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7566506
Report Number0003015876-2018-00864
Device Sequence Number16816151
Product Code MKJ
UDI-Device Identifier00883873911723
UDI-Public00883873911723
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001247
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Initial Date Received by Manufacturer 05/08/2018
Supplement Date Received by Manufacturer11/12/2018
Initial Report FDA Received Date06/04/2018
Supplement Report FDA Received Date11/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-