• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE Back to Search Results
Catalog Number 8523.210
Medical Device Problem Code Device Operational Issue (2914)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/11/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint was reviewed during the retrospective review.The investigation was done on the returned cutter.The cutter was visually inspected, no deviations where found.The cutter was working according to specifications.Only when the aspiration was turned off bubbles appeared.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4)).Details of this activity were discussed with cdrh office of compliance ((b)(6)) during a tele-conference on (b)(6) 2017.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
Air came out from the tip.Sample is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE
Common Device Name
DISPOSABLE HIGH SPEED TDC CUTTER 25 G / 8000 CPM DORC CONTINUUM RANGE
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7566563
Report Number1222074-2018-00168
Device Sequence Number2012339
Product Code HQE
UDI-Device Identifier08717872020808
UDI-Public08717872020808
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 05/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number8523.210
Device Lot Number2000371911
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2016
Initial Report FDA Received Date06/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-