| Model Number |
461438 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that a 6.5 fr polyurethane feeding tube was secured in place.The purple cap/top accessed to hook up the syringe to check placement and purple top disconnected.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Two decontaminated samples were received at the plant for the investigation.The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Upon a visual evaluation of the sample, the defect was confirmed; the separation of the purple connector can be observed.A root cause could not be determined.As part of continuous improvements, a corrective action has been opened to implement effective solutions to prevent the reoccurrence of the reported condition.This complaint will be used for tracking and trending purposes.If information is provided in the future, a supplemental report will be issued.
|
| |
|