| Model Number |
2658-01-0002 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The light could not be requested back because it had been in contact with the patient.Pictures were provided.The lens appears to be intact leading to the investigators to believe that there was an issue with the adhesive.The operators who assemble the lights were interviewed and the dhr was reviewed.There was nothing that indicated anything unusual.The evaluation is inconclusive.In the eleven years these surgical light models have been on the market, this is the first incident where the lens came off.
|
| |
|
Event or Problem Description
|
|
The lens from the surgical light fell off while in use on a patient.The lens was able to be retrieved.There was no injury, delay in surgery, or any additional procedures required.
|
| |
|