• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC WIDEBAND L 100BX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC WIDEBAND L 100BX Back to Search Results
Model Number 36104
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the adhesive on the male external catheters were too strong, which made it difficult to remove from the skin.The patient allegedly used a warm compress to help remove the male external catheter, however it took a long time for it to be removed.
 
Event or Problem Description
It was reported that the adhesive on the male external catheters were too strong, which made it difficult to remove from the skin.The patient allegedly used a warm compress to help remove the male external catheter, however it took a long time for it to be removed.
 
Additional Manufacturer Narrative
The reported event was unconfirmed.Visual evaluation of the samples noted no obvious defects.An adhesive peel test was performed and the samples were cut (0.7095", 0.7265") to the required width (0.70" +/- 0.05 ").The adhesive peel strength (0.6689lbf, 0.6546lbf) was found to be low and out of specification (0.80-2.80lbf) for the samples tested.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC WIDEBAND L 100BX
Common Device Name
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7566767
Report Number1018233-2018-02031
Device Sequence Number1523786
Product Code NNX
UDI-Device Identifier00801741071003
UDI-Public(01)00801741071003
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/28/2022
Device Model Number36104
Device Catalogue Number36104
Device Lot NumberJUBT0469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer08/09/2018
Initial Report FDA Received Date06/04/2018
Supplement Report FDA Received Date08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-