STRYKER GMBH OSTEOSYNTHESIS COMPRESSION STAPLE EASYCLIP 10X15X13MM; SINGLE/MULTIPLE COMPONENT METALLIC BONE FIXATION APPLIANCES AND ACCESSORIES. Back to Search Results
Catalog Number EZB101513
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nerve Damage (1979)
Date of Event 04/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event that osteosynthesis compression staple easyclip 10x15x13mm was alleged of 'wrong device used' could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.Based on investigation, the root cause was attributed to be user related.The failure was caused by an off-label use.It was reported that this easyclip staple was implanted in the femoral area, when this device is dedicated to hand and foot surgeries.Please note that the operative technique (ec-st-1-en_ easyclip optech_2015-7180) and the instruction for use (v15082 rev aa easyclip_memoclip_4fusion) read: "indications the easyclip staples are indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.'' [original statement(s)] therefore this case is classified as user related.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.Remains implanted.
 
Event or Problem Description
The surgeon reported : "last tuesday i operated a patient for a thp via dsa.The procedure went smoothly, following the tips from the instructional guide from stryker.Postoperatively the patient developed an n femoralis outage." the customer asks for advise from an expert to avoid this in the future.Update per additional information received (b)(6) 2018 - doctor did a primary hip with a tritanium cup and a exeter femoral stem.He used for the approach the dsa technique , ( direct superior approach ) which is promoted by us since a couple of years.After the operation the patient had a femoralis drop , what is something like a paralysis of the sciatic nerve.There was no revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEOSYNTHESIS COMPRESSION STAPLE EASYCLIP 10X15X13MM
Common Device Name
SINGLE/MULTIPLE COMPONENT METALLIC BONE FIXATION APPLIANCES AND ACCESSORIES.
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7567127
Report Number0008031020-2018-00416
Device Sequence Number19504914
Product Code JDR
UDI-Device Identifier07613252259367
UDI-Public07613252259367
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K070031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/30/2022
Device Catalogue NumberEZB101513
Device Lot NumberL55424
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2018
Initial Report FDA Received Date06/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
-
-