Model Number 15
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that their device had a service indicator present and an event code in its memory which indicates that the aed function of the device may be inoperable and, as a result, the unit may not be able to analyze a patient's rhythm and provide defibrillation therapy while in aed mode, if necessary.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control replaced the therapy pcb assembly and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio-control further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was an electrically leaky capacitor, designator c160.
 
Event or Problem Description
The customer contacted physio-control to report that their device had a service indicator present and an event code in its memory which indicates that the aed function of the device may be inoperable and, as a result, the unit may not be able to analyze a patient's rhythm and provide defibrillation therapy while in aed mode, if necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7567129
Report Number0003015876-2018-00867
Device Sequence Number1760153
Product Code MKJ
UDI-Device Identifier00883873911730
UDI-Public00883873911730
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001246
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Initial Date Received by Manufacturer 05/10/2018
Supplement Date Received by Manufacturer09/25/2018
Initial Report FDA Received Date06/04/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-