| Model Number |
15 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had a service indicator present and an event code in its memory which indicates that the aed function of the device may be inoperable and, as a result, the unit may not be able to analyze a patient's rhythm and provide defibrillation therapy while in aed mode, if necessary.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control replaced the therapy pcb assembly and completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio-control further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was an electrically leaky capacitor, designator c160.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had a service indicator present and an event code in its memory which indicates that the aed function of the device may be inoperable and, as a result, the unit may not be able to analyze a patient's rhythm and provide defibrillation therapy while in aed mode, if necessary.There was no patient use associated with the reported event.
|
| |
|