| Model Number |
EXPRESS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device did not power on when prompted.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio-control determined that the cause of the reported issue was that the device's lid latch was pushed off the case shaft and, as a result, the on/off switch could not be activated in order to power the device on.The customer was provided with a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device did not power on when prompted.There was no patient use associated with the reported event.
|
| |
|