| Catalog Number |
365987 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
09/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode with the incident lot was observed.Additionally, retention samples were selected for evaluation, and the customer's indicated failure mode was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that bd microtainer® pst¿ tubes with lh lithium heparin had a sticky substance inside the tube with a stringy substance attached to the cap.
|
| |
|
Additional Manufacturer Narrative
|
|
Medical device manufacturer corrected to (b)(4).Manufacturing device updated to (b)(4).
|
| |
|
Search Alerts/Recalls
|