| Catalog Number |
11-4646 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation in progress; no conclusion can be made at this time.Upon completion of the evaluation, a follow up supplemental report shall be submitted.
|
| |
|
Event or Problem Description
|
|
During placement of a pedicle screw for lumbar fixation, the surgeon experienced difficulty reducing the rod at the l4-l5 level.Subsequently, the surgeon noticed the screw was separated.The screw was removed and replaced intraoperatively.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated and the reported event was confirmed.The returned device displayed visual indications of use in a manner that exceeded the strength of the assembly.Units from inventory were inspected and were found to meet specifications.A review of the dhr found the product met release requirements.
|
| |
|