• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926024250
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a coronary vessel.A 2.50 x 24 synergy ii stent was advanced to treat the lesion.However, upon advancing the device, it was noticed under fluoroscopy that the stent did not look right.The device was removed and it was noted that the stent struts were damaged.The procedure was completed with a different device.No patient complications were reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: stent delivery system was returned for analysis.A visual examination of the stent found that stent struts from the proximal to the mid-section of the stent were damaged, stent struts were stretched, pushed and bunched in a distal direction around the mid-section of the stent.The undamaged distal section of the crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded on the distal balloon cones and were not subjected to positive pressure.Stent crimp markings were present on the exposed proximal balloon body.The tip was visually and microscopically examined and no signs of damage were noted.A visual and tactile examination of the hypotube found multiple kinks along several locations of the hypotube shaft.This type of damage is consistent with excessive force being applied on the delivery system.A visual and tactile examination of the outer and the inner lumen and mid-shaft section found no issues with the extrusion shaft.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a coronary vessel.A 2.50 x 24 synergy ii stent was advanced to treat the lesion.However, upon advancing the device, it was noticed under fluoroscopy that the stent did not look right.The device was removed and it was noted that the stent struts were damaged.The procedure was completed with a different device.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7567603
Report Number2134265-2018-04954
Device Sequence Number14110769
Product Code NIQ
UDI-Device Identifier08714729840213
UDI-Public08714729840213
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2019
Device Model NumberH7493926024250
Device Catalogue Number39260-2425
Device Lot Number0021717754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Initial Date Received by Manufacturer 05/14/2018
Supplement Date Received by Manufacturer06/12/2018
Initial Report FDA Received Date06/04/2018
Supplement Report FDA Received Date06/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No