• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. AZUR DETACHABLE FRAMING COIL; EMBOLIZATION COIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. AZUR DETACHABLE FRAMING COIL; EMBOLIZATION COIL Back to Search Results
Model Number 45-652050
Medical Device Problem Codes Detachment Of Device Component (1104); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history record indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was not returned to the manufacturer for analysis; therefore, a product evaluation could not be performed.The root cause is unknown.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report or the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
 
Event or Problem Description
It was reported that an embolization coil got stuck in the catheter and detached.The coil was removed in its entirety along with the catheter.There was no reported intervention or patient injury.The patient was reported to be great.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR DETACHABLE FRAMING COIL
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7567912
Report Number2032493-2018-00103
Device Sequence Number14122755
Product Code KRD
UDI-Device Identifier00810170016158
UDI-Public(01)00810170016158(11)180302(17)230228(10)18030253R
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111307
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/28/2023
Device Model Number45-652050
Device Catalogue Number45-652050
Device Lot Number18030253R
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/07/2018
Initial Report FDA Received Date06/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-