MEDTRONIC IRELAND ENDURANT II EXTENSION CUFF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
ETTF3636C70E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
05/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An anuerx stent graft system was implanted in a patient for the endovascular treatment of an unknown size abdominal aortic aneurysm.An endurant ii cuff was also implanted approximately 6 years post the index procedure for a suspected stent graft migration and type ia endoleak.It was reported that the patient presented emergently approximately 12 years post the index procedure with abdominal pain.A ct showed a type iii endoleak between the cuff and bifurcated stent graft.The physician reported the cause of the event to be related to patient anatomy and reported there was a morphology change due to angulation between the neck and existing bifurcated stent graft.It was reported that the current diameter of the aorta at the renal artery was 35 mm, the current length of the proximal aortic neck was 5mm, and the degree of angulation was 50.The patient was treated with an endurant ii abdominal tube, an endurant ii limb extension, seven endoanchors and a reliant balloon.A non-mdt plug was also implanted in the patient and a fem-fem bypass carried out.Either a type ia or type ii endoleak was noted on the final angiogram.The event was reported to have been resolved.No additional clinical sequelae were reported and the patient is being monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Film evaluation summary: the exact cause of the events could not be determined from the limited films provided.The anatomy at implant is unknown, and earlier post-implant films were not available for review.Two post-implant ct¿s slices and a single still angiogram from an unspecified study date were returned.The films confirmed that the aneurx bifurcate had migrated into the sac and was positioned >8cm below the renal arteries.An endurant aortic extension was seen implanted just below the renal arteries, and there was a 1 ¿ 2 cm gap between the distal end of the extension and the bifurcate with contrast seen at the separation from a type iiia junctional endoleak.Migration of the 28mm diameter aneurx bifurcate appeared to have been caused by disease progression with dilatation of the neck diameter to ~35mm, as well as possible aneurysm morphology changes as seen with the current severe neck and iliac limb angulation.The exact cause of the component detachment leading to the type iiib endoleak is unknown.Films just prior to the cuff implant were not available and the amount of initial device overlap is unknown.It is likely that morphology changes as well as severe diameter oversizing between the 36mm aortic extension and bifurcate may have also contributed to the type iii junctional separation within the unsupported aneurysm sac.The exact cause of the reported endoleak noted on the final angiogram after implant of an abdominal tube, limb extension, and multiple endoanchors also could not be determined.Films were not available.It is possible that implanting within the large diameter (35mm) and short >10mm) length neck which was also angulated may have all contributed to a potential proximal type i endoleak.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|