The device was not returned by the user facility, therefore; the reported failure could not be verified and a root cause cannot be conclusively determined.A review of the manufacturing documents verified the device was produced per current and approved procedures and material specifications.Non-conformances regarding the product's identity, quality, safety, effectiveness or performance were not identified at the time of manufacture.A review of this lot number showed no other complaint for this lot product and event description.A two-year review of complaint history revealed 17 other similar complaints, with 19 devices, for this product family and failure mode.In that same timeframe, (b)(4) units of this product family have been sold worldwide, making the rate of occurrence of this failure (b)(4) percent, if all complaints were confirmed.The instructions for use advises the user of the following warnings and contraindications.Inspect instruments prior to use to ensure they are in good physical condition and function properly.There should be no loose, broken or misalignment parts.Exercise care in the use of these devices to minimize side or bending loads.Do not use excessive force on instruments during use to prevent damage or breakage.Inspect instruments after use to ensure they have not been damaged.Avoid lateral loading while inserting y-knot anchors.Maintain proper alignment during insertion of anchors and disengagement of drivers.This issue will continue to be monitored through the complaint system to assure patient safety.
|
|
The conmed sales representative reported on behalf of the user facility that during a distal bicep repair, the surgeon inserted two y1301 y-knot flex anchors into the bicipital tuberosity after previously drilling the holes with the standard guide.After inserting the anchors, while removing the driver and pulling on the suture to seat the anchor, the nurse noticed that one of the forks on both anchor's drivers had broken off.There were no issues reported with the anchors, as they were correctly seated with the suture.The surgeon conducted a search for the forks, but was unable to locate the fork tips.The surgical team concluded the broken piece was washed and suctioned.The procedure was completed with a 2 minute surgical delay and no reported patient injury.This report is being raised on the basis of the location of a broken piece of a non implantable device being unknown.The other reported y1301 y-knot flex is being reported under mdr 1017294-2018-00075.
|