Model Number 5100C-PA
Medical Device Problem Codes False Reading From Device Non-Compliance (1228); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure, the device had a reading issue: the value of around 40 which appeared on the monitor did not match with the curve around 80.The customer reported no patient harm.
 
Additional Manufacturer Narrative
Evaluation summary: one device was evaluated.Service center technician reported that there was no failure was found.The investigation found the device to function normally.The reported product was not related to a reported complaint event.There was no fault with the pre amp unit.The failure was attributed to the monitor which was also returned for investigation.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure, the device had reading issue, the value of around 40 which appeared on the monitor did not match with the curve around 80.This was also tested a few days later, same thing happened.The range difference was not as big as in the patient's case.The customer reported no patient injury.
 
Additional Manufacturer Narrative
Correction: d3 additional information: d8, g1, g3, h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway 20023 3
EI  200233
Manufacturer (Section G)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway 20023 3
EI   200233
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, MA 55112
7635265677
MDR Report Key7568628
Report Number8020893-2018-00248
Device Sequence Number11311226
Product Code MUD
UDI-Device Identifier10884521145573
UDI-Public10884521145573
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K082327
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5100C-PA
Device Catalogue Number5100C-PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2018
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 05/14/2018
Supplement Date Received by Manufacturer05/14/2018
08/08/2024
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date12/26/2018
08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
Patient Weight64 KG
-
-