| Model Number |
MMT-723RNAB |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
05/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump comprised force sensor system error alarm.The customer¿s blood glucose level was 301 mg/dl at the time of the incident.Alarm occurred during the rewind.The product will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test.Insulin pump received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or verify compromised forced sensor system alarm due to motor error alarm.
|
| |
|