| Catalog Number |
8000.COM02 |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Codes |
Intraocular Pressure, Delayed, Uncontrolled (1936); No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint was reviewed during the retrospective review.Investigation determined that the pump is working according specifications, our suspicion is that there was not enough fluid in the bottle or a blocked tube, which caused no infusion.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer ((b)(6).).Details of this activity were discussed with (b)(6) during a tele-conference on (b)(6) 2017.All available information has been disclosed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
At our customer there are infusion problems on eva.The surgeon still hears the sound of the drops/drop noise, but there is no infusion, why the eyes collapse.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint was reviewed during the retrospective review.Investigation determined that the pump is working according specifications, our suspicion is that there was not enough fluid in the bottle or a blocked tube, which caused no infusion.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre b.V.).Details of this activity were discussed with cdrh office of compliance (b)(4)) during a tele-conference on (b)(6) 2017.All available information has been disclosed.This complaint has now been closed.
|
| |
|
Event or Problem Description
|
|
At our customer there are infusion problems on eva.The surgeon still hears the sound of the drops/drop noise, but there is no infusion, why the eyes collapse.
|
| |
|