Model Number R SERIES
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during functional testing, the device displayed "system fault 36" and "system fault 37" messages.Complainant did not indicate that there was any patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The reported malfunction was observed during review of the activity logs.However, after extensive testing the reported problem could not be duplicated.The device's analog board was replaced as a precaution.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7568863
Report Number1220908-2018-01535
Device Sequence Number1317699
Product Code MKJ
UDI-Device Identifier00847946002350
UDI-Public00847946002350
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K060559/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Initial Date Received by Manufacturer 05/15/2018
Supplement Date Received by Manufacturer05/15/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-