JOHNSON & JOHNSON SURGICAL VISION TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 03/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that on day one post-op due to anterior capsule - posterior capsule rupture, the intraocular lens (iol) didn't stay centered as the capsular bag wasn¿t stable as the surgeon expected.Reportedly, the retinal doctor did a vitrectomy and explanted the intraocular lens (model zlb00, +18.5 diopter) in a secondary surgical procedure from the patient¿s operative eye.A competitor monofocal three-piece iol was used as a replacement lens.The patient was reported doing excellently when discharge.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 7/20/2018.Device returned to manufacturer? yes.Device evaluation: the device was received at the manufacturing site for investigation.Visual inspection shows the product was cut into two pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Historical data analysis: the search in complaint history revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION
santa ana CA
MDR Report Key7568919
Report Number3011852734-2018-00095
Device Sequence Number10807274
Product Code MFK
UDI-Device Identifier05050474560642
UDI-Public(01)05050474560642(17)210611
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/05/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/11/2021
Device Model NumberZLB00
Device Catalogue NumberZLB00U0185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/07/2018
Supplement Date Received by Manufacturer08/14/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Date Report Sent to FDA01/01/2005
-
-