Catalog Number 8000.COM02
Medical Device Problem Code Device Operational Issue (2914)
Health Effect - Clinical Code Intraocular Pressure, Delayed, Uncontrolled (1936)
Date of Event 06/14/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint was reviewed during the retrospective review.Investigation determined an unstable reducer and defective pcb.After replacing this parts and readjusting the air, the vfie module is working according specifications.The complaint is confirmed.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer ((b)(4).Details of this activity were discussed with (b)(6) during a tele-conference on (b)(6) 2017.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
A report from (b)(6) has been received and report a soft eye issue when switching from phaco to vitrectomy in 25 combined prog.At our customer there are infusion problems on eva.The surgeon still hears the sound of the drops/drop noise, but there is no infusion, why the eyes collapse.There is no information on intervention provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR)
Common Device Name
EVA, COMBINED MACHINE INCLUDING LASER (DORC CONNECTOR)
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7568989
Report Number1222074-2018-00171
Device Sequence Number11311225
Product Code HQC
UDI-Device Identifier08717872019659
UDI-Public08717872019659
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 06/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number8000.COM02
Device Lot Number2013000104
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2016
Initial Report FDA Received Date06/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-