| Model Number |
N/A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Blood Loss (2597); No Code Available (3191)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.Patient code: other: surgery; removed device fragment from patient.The actual device was not returned to the manufacturing facility for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported that a 65cm destination device had pulled apart in a patient.It was reported that the vessels were very calcified and there was a lot of resistance.There was a piece left in the patient but the piece was removed.The estimated blood loss was less than 250cc.The patient's condition is ok and the procedure outcome was reported to be ok.Additional information was received on 5/31/2018: the patient is doing fine and was able to leave the hospital after a few days.Medical intervention was performed to remove the broken piece of the destination.Additional information was received on 6/1/2018: the patient was taken to surgery to have the broke piece of the destination removed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide additional information, to provide the completed investigation results and the medwatch.With no device return, the exact cause of the reported event cannot be definitively determined based on the available information.Mw5077420.
|
| |
|
Event or Problem Description
|
|
Additional information was received on 6/5/18: the patient had previous surgeries in the groin area.Scar tissue was present.The physician started with a 5fr diagnostic sheath and then switched to the 65cm destination.The destination was tight on insertion due to the scar tissue.Vessel characteristics: moderate tortuosity.No calcium present in the area of the destination.The access site was right side contralateral and the procedure was a left lower extremity angioplasty.The procedure was performed.At the end of the case, the physician was removing the destination over a stiff guidewire.At first there was a little resistance, then it became very easy to pull back the sheath.The physician noticed that the destination was unraveling.The physician cut the destination at the entry site.Surgical intervention was performed.A cut down was performed on both groins and all of the destination fragments were removed from the patient.The patient recovered well and was discharged 2 days post procedure.
|
| |
|