It was reported the an advance 18 lp low profile balloon catheter ruptured (refer to this report).Additionally, the physician reported three other occurrences where advance 18 lp low profile balloon catheters (lots unknown), have ruptured (refer to medwatch: 1820334-2018-01601).No further information was provided.No adverse events have been reported.Additional information regarding event details, patient anatomy and outcome has been requested, but is not available at this time.
|
|
The customer confirmed that, during an "angio" [sic] of the superficial femoral artery, after access was obtained and the balloon was advanced to the site of inflation, the inflated balloon ruptured longitudinally at a pressure of 12 atmospheres (atm).The customer confirmed that the vessel was neither tortuous nor particularly calcified, although the calcification was characterized as minimal.The procedure was completed successfully using another balloon device.Visipaque contrast was employed with saline, with a 40/60 ratio employed.There was no angulation.Concomitant products: v18 wire (boston scientific), encore inflation handle (boston scientific), 5 french sheath (terumo).Investigation - evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), quality control, specifications, and a visual inspection of the complaint device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a longitudinal balloon rupture.The rupture was measured to begin a distance of 0.6 cm from the proximal marker band, and it extended to a length of 2.9 cm from the distal tip of the device.Both marker bands were present, as was the distal tip of the catheter.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed one nonconforming event which could contribute to this failure mode.However, due to the individual nature of the product manufacturing and 100% verification of various defects in final quality control prior to shipping, there is no indication that there is nonconforming product in the field.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
|