Additional information received on this case reported that the physician performed a revision to fix the occluded limb.An endurant ii 25x25x49 cuff was placed in the existing main body stent graft and an extension limb (16x13x124mm) was implanted in the right ipsi side to realign graft.Another non mdt vbx device was implanted in the right limb to reinforce area and as that device appeared kinked, another device was extended distally to the external artery.Ballooning was performed with reliant in the proximal main body and iliac limb and kissing balloons were used in the distal aorta.Post angiogram showed both limbs patent with good flow bilateral and the patient had good pulses bilaterally post procedure films review summary: the exact cause of the reported right limb occlusion could not be determined from the pre-implant films provided.Films at implant and post-implant ct¿s were not provided, and the in-vivo configuration of the implanted stent grafts and the reported occlusion event could not be assessed.Review of a pre-implant 3d/ct film report revealed that the patient had a proximal neck that ranged in diameter from 20 ¿ 21mm along a 3cm length neck that was essentially straight and free from calcification and thrombus.The distal aortic diameter measured 21mm with calcification noted.The right common iliac artery was non-tortuous, measuring from 10 ¿ 11.5mm in diameter, and the right external iliac artery measured 10mm in diameter near its origin.It is possible that this occlusion event was anatomy related; implanting within a small calcified distal aorta that may have compressed the right limb, leading to thrombus formation.Other possible occlusion contributors (which are currently unknown) include verification of the diameter of the implanted contra limb(s) and possible ipsi limb extension, amount of limb/vessel oversizing, assessment of the limb flow lumen along the entire length, location of any overlapping limbs, and if the right limb may have been extended into the right external iliac artery.This occlusion may also have been caused by a patient condition: poor distal runoff, an unknown coagulopathy that is now presenting, or a previous history of pad.If information is provided in the future, a supplemental report will be issued.
|