| Catalog Number |
1020-5204 |
| Medical Device Problem Code |
Delamination (2904)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Color marking from the rasps crumble off.Replacement was available.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding crack/fracture involving an accolade broach was reported.The event was confirmed by product inspection.Conclusions: the reported event was confirmed as per material analysis performed on the returned product which concluded that the neck indicator fractured in multiple overload events.No identifiable material or manufacturing discrepancies were observed on the surfaces examined.No further investigation is required at this time.If the additional information is received, this investigation will be reopened and re-evaluated.
|
| |
|
Event or Problem Description
|
|
Color marking from the rasps crumble off.Replacement was available.
|
| |
|
Search Alerts/Recalls
|