Model Number 9500-44
Device Problems Imprecision (1307); Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted into the abdomen on (b)(6) 2018.No additional event or patient information is available.Data was provided.Review of the data log confirmed the report of an inaccuracy.Additionally, it was determined via data that calibrations were entered during periods of rapid rate of change.The root cause was determined to be improper calibration during a rapid rise or fall.Labeling indicates: do not calibrate if your blood glucose is changing at a significant rate, typically more than 2 mg/dl per minute.Do not calibrate when your receiver screen is showing the rising or falling single arrow or double arrow, which indicates that your blood glucose is rapidly rising or falling.Calibrating during rapid rise or fall of blood glucose may affect sensor accuracy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7569500
MDR Text Key110506088
Report Number3004753838-2018-060534
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9500-44
Device Catalogue NumberSTS-GF-007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age15 YR
Patient Weight64
-
-