Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, there was an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter.The sensor was inserted into the abdomen on (b)(6) 2018.No additional event or patient information is available.Data was provided.Review of the data log confirmed the report of an inaccuracy.Additionally, it was determined via data that calibrations were entered during periods of rapid rate of change.The root cause was determined to be improper calibration during a rapid rise or fall.Labeling indicates: do not calibrate if your blood glucose is changing at a significant rate, typically more than 2 mg/dl per minute.Do not calibrate when your receiver screen is showing the rising or falling single arrow or double arrow, which indicates that your blood glucose is rapidly rising or falling.Calibrating during rapid rise or fall of blood glucose may affect sensor accuracy.
|