• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC VENACURE NEVERTOUCH FIBER; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC VENACURE NEVERTOUCH FIBER; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number H787114030010
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.The firm is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on may 08.2018: the treating physician withdrew the fiber and the sheath from the patient as one unit.When disassembling in order to flush and clean it, the tip of the fiber appeared to be stuck inside of the sheath.It was noted the tip had melted to the inside of the sheath.The device was set aside and the procedure was successfully completed witha new of the same device.There was no harm or injury to the patient due to this event.It was reported the disposable device is available for return to the manufacturer for a device evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was a nevertouch fiber.A visual review of the fiber noted that the gold tip tube had detached from the fiber and was stuck in the end of the tre-sheath.The returned devices met all manufacturing dimensional specifications for this product.The customer's reported complaint description of the "tip of the fiber lodged the sheath tip" is confirmed; the sheath was not melted.Evaluation of the returned sample confirmed the fiber was manufactured correctly, as evidenced by proper crimp marks on the tube tip and adhesive present.This is also supported by the fact that there were no issues reported with the use of the fiber/sheath during the first vein ablation.A definitive root cause cannot be determined at this time, although a possible contributing factor could be due to a buildup of biological matter (dried blood) in the tre-sheath that occurred during the ablation.This buildup prevented fiber tip from being pulled back into the sheath at the end of the first vein ablation.The end user then pulled back on the fiber which resulted in the tip detaching from the fiber shaft.A review of the angiodynamics' device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number, contains the following statement "precaution: prior to and during use, avoid damaging the fiber by striking, stressing, or excessive bending.Do not coil the fiber tighter than a diameter of 20cm.Clinical safety and effectiveness data is not available for other fiber tip designs and diameters." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENACURE NEVERTOUCH FIBER
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS, INC
603 queensbury avenue
queensbury NY 12804
MDR Report Key7569888
Report Number1319211-2018-00059
Device Sequence Number2678093
Product Code GEX
UDI-Device IdentifierH787114030010
UDI-PublicH787114030010
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K162914
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue NumberH787114030010
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-