MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS HIGH FLEXION INSERT SZ 1-2 9MM; PRSTHSS,KNE,PTLLFMRTBL,SMCNSTRND,CMNTD,PLYMRMTLPLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS HIGH FLEXION INSERT SZ 1-2 9MM; PRSTHSS,KNE,PTLLFMRTBL,SMCNSTRND,CMNTD,PLYMRMTLPLYMR Back to Search Results
Catalog Number 71421501
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Joint Dislocation (2374)
Date of Event 05/23/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to dislocation.
 
Additional Manufacturer Narrative
The associated genesis ii ps high flexion insert was returned and evaluated.A clinical evaluation noted that without supporting clinical/medical documents a thorough investigation cannot be performed.A lab analysis conducted during this investigation noted that lock detail deformation and scratches are observed on the inferior surface of the insert.The exact cause could not be determined from the investigation of the device.There were no observations of material or manufacturing deviations in the course of this investigation.Our investigation including a review of complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing or product specifications for the listed batch did not reveal abnormalities that could have contributed to the reported issue.No further investigation warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HIGH FLEXION INSERT SZ 1-2 9MM
Common Device Name
PRSTHSS,KNE,PTLLFMRTBL,SMCNSTRND,CMNTD,PLYMRMTLPLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7569904
Report Number1020279-2018-01063
Device Sequence Number16584060
Product Code JWH
UDI-Device Identifier03596010492388
UDI-Public03596010492388
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K032295
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71421501
Device Lot Number16JT75456
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-