STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL HA SOLID BACK 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 540-11-56F
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 05/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding dislocation involving a trident shell was reported.The event was not confirmed.Method and results: product evaluation and results: not performed as the device was not returned.Medical records received and evaluation: not performed as no medical records were returned for review.Product history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that patient's right hip was revised after recurrent anterior subluxation.Intra-operatively, hematoma was addressed.Patient was revised to a trident tritanium shell with screws, adm/mdm liner construct, and ceramic head.
 
Event or Problem Description
It was reported that patient's right hip was revised after recurrent anterior subluxation.Intra-operatively, hematoma was addressed.Patient was revised to a trident tritanium shell with screws, adm/mdm liner construct, and ceramic head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL HA SOLID BACK 56MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7570135
Report Number0002249697-2018-01714
Device Sequence Number2629660
Product Code LPH
UDI-Device Identifier04546540237859
UDI-Public04546540237859
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryGB
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number540-11-56F
Device Lot Number62811104
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/11/2018
Supplement Date Received by Manufacturer06/06/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient Weight86
-
-