It was reported that the patient¿s generator settings spontaneously changed to 0ma and the physician reports that the patient¿s battery was only 25% and the device was implanted less than one year ago.Data from the patient's device was received and reviewed.It was seen that the device was set to vbat < eos threshold when the device was interrogated on (b)(6) 2018.All outputs at this time were set to 0 ma including the magnet output.Lead impedance was within normal limits.The device was then programmed back on to desired settings.From reviewing the data, the generator is functioning properly and thus the only explanation for why the generator was at settings of 0ma upon interrogation is that the generator was likely in contact with some form of electrocautery on or around (b)(6) 2018.This may have been caused by event non-invasive procedure where possible electrocautery or radio frequency tools were used.As far as the generator¿s performance right now, all impedance values are within normal limits and the voltage of the battery (battery longevity) appears normal and was not significantly affected by this electrocautery use.No additional or relevant information has been received to date.
|
Product analysis for the generator was completed and approved.The pulse generator was opened.A visual assessment on the pcba showed a reflective substance on the trimmed edge of the pcba (tab removed).In addition a white residue was observed surrounding all leads of the pcba processor.No other visual anomalies were identified.An intensified-follow-up-indicator warning message was verified in the analysis lab and found to be due to the pulse generator remaining ¿on¿ post explant.The device was opened and analyzed.During investigation, a white residue was observed surrounding the leads of the pcba processor.In addition, a reflective substance was observed on the trimmed edge of the pcba (tab removed), suggesting potential for an electrical path (resistive path), between the copper edges on the trimmed edge of the pcba.It is unclear if these findings may have been a contributing factor for the premature end of life indicator allegation.With the case removed and the battery still attached to the pcba, the series resistor measured 0.35mv, indicating no high current draw state and the battery measured 2.87 volts.
|