ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.
 
Event or Problem Description
On (b)(6) 2018, a reporter contacted animas alleging that the pump was alarming.This complaint is being reported because the alleged issue could cause a delay in insulin delivery.There was no serious injury associated with this complaint.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 19-dec-2019 with the following findings: during investigation, the pump is unable to power on due to impact damage to display and pcb components.Unable to investigate pump is alarming complaint due to impact damage.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Event or Problem Description
Hold by yd, 1/7/2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7570516
Report Number2531779-2018-10246
Device Sequence Number1262209
Product Code OYC
UDI-Device Identifier10840406100341
UDI-Public10840406100341
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Was the Report Sent to FDA? Yes
Device Age17 MO
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date12/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age19 YR
-
-