Model Number 1425US
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Concomitant medical products: 1103 vad implanted: (b)(6) 2017, dtba1q1 icd implanted: (b)(6) 2016, 429888 lead implanted: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the controller ac adapter did not provide power and the green light did not illuminate.The controller ac adapter will be exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the controller ac adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned device revealed that the unit passed visual inspection and functional testing.As a result the reported event could not be confirmed.Based on the investigation conducted, possible causes of the reported event can be attributed to multiple factors, including but not limited to, a marginal connection between the controller ac adapter and the wall power outlet, and/or a marginal connection between the controller ac adapter and the controller.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The controller ac adapter was exchanged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7570537
Report Number3007042319-2018-02238
Device Sequence Number1759246
Product Code DSQ
UDI-Device Identifier00888707000307
UDI-Public00888707000307
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/06/2018
Device Model Number1425US
Device Catalogue Number1425US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Initial Date Received by Manufacturer 05/09/2018
Supplement Date Received by Manufacturer08/22/2018
Initial Report FDA Received Date06/05/2018
Supplement Report FDA Received Date08/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694765 LEAD, 5076-52 LEAD
Patient Age49 YR
Patient Weight62
-
-