BECTON DICKINSON MEDICAL SYSTEMS 10 ML BD¿ PRE-FILLED NORMAL SALINE SYRINGE, IN 10 ML SYRINGE; PREFILLED SALINE FLUSH SYRINGE
|
Back to Search Results |
|
Catalog Number 306500 |
Device Problem
Nonstandard Device (1420)
|
Patient Problem
Unspecified Infection (1930)
|
Event Date 04/29/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that a patient was admitted to a hospital on (b)(6) 2018 through (b)(6) 2018 because she developed two infections.The patient indicated that the infections were associated with the use of a 10 ml bd¿ pre-filled normal saline syringe, in 10 ml syringe.Information about specific treatment provided in the hospital was not provided.Out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.
|
|
Manufacturer Narrative
|
Investigation summary: lot number: 714411a for product code: 306500 was provided for evaluation by our quality engineer team.Upon reviewing the production history for the provided lot number, no deviations or non-conformances were identified during the manufacturing process.A corrective and preventive action plan was initiated to further investigate and monitor this issue.Prior to the report of infection evaluated in mps-18-1248-sa, there was no trend for infection cases reported for franklin product.This complaint is part of a new trend which began in april 2018.There were no deviations, non-conformances, or out of specification conditions noted during the manufacture of lot: 714411a 100% of retained samples for the lot: (b)(4) units) were visually inspected.No growth was seen (i.E., the solution was clear).Lot: 74411a was manufactured between lots: 705311b and 716192n, both of which exhibited no growth during the confirmatory sterility testing.This, in combination with the daily environmental monitoring and original sterility testing for the lot release, provides confidence in the sterility of the complaint lot.A review of all lot sterility testing performed for product released between april 2015 & june 2018 confirmed that no organism growth was identified for any lots released during the shelf-life of this product.This in combination with the daily environmental monitoring and sterility testing provides confidence in the sterility of the reported lot number.The root cause analysis of the reported infection cases under capa: 350041 has not identified a direct causation between the infections and the bd franklin product.
|
|
Event Description
|
It was reported that a patient was admitted to a hospital on (b)(6) 2018 through on (b)(6) 2018 because she developed two infections.The patient indicated that the infections were associated with the use of a 10 ml bd¿ pre-filled normal saline syringe, in 10 ml syringe.Information about specific treatment provided in the hospital was not provided.Out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.
|
|
Search Alerts/Recalls
|
|
|