• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS 10 ML BD¿ PRE-FILLED NORMAL SALINE SYRINGE, IN 10 ML SYRINGE; PREFILLED SALINE FLUSH SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS 10 ML BD¿ PRE-FILLED NORMAL SALINE SYRINGE, IN 10 ML SYRINGE; PREFILLED SALINE FLUSH SYRINGE Back to Search Results
Catalog Number 306500
Device Problem Nonstandard Device (1420)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 05/14/2014
Event Type  Injury  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that a patient became sick in (b)(6) and had kidney surgery.The patient associated his incident with a 10 ml bd¿ pre-filled normal saline syringe, in 10 ml syringe.The patient declined to provided further information.*out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.
 
Manufacturer Narrative
Investigation summary: lot number 717911c for product code 306500 was provided for evaluation by our quality engineer team.Upon reviewing the production history for the provided lot number, no deviations or non-conformances were identified during the manufacturing process.A corrective and preventive action plan was initiated to further investigate and monitor this issue.One-hundred and twenty retained samples for the lot number provided were visually inspected and no abnormalities were observed in regards to the solution.A review of all lot sterility testing performed for product released between april 2015 and june 2018 confirmed that no organism growth was identified for any lots released during the shelf-life of this product.Multiple samples produced before and after the provided lot number were sent for sterility testing and confirmed that no microbial growth was exhibited after incubation.This in combination with the daily environmental monitoring and sterility testing provides confidence in the sterility of the reported lot number.A direct causation between the reported reactions and infections and the bd franklin product has not been identified.Investigation conclusion: the root cause analysis of the reported infection cases under capa (b)(4) has not identified a direct causation between the infections and the bd franklin product.
 
Event Description
It was reported that a patient became sick in january and had kidney surgery.The patient associated his incident with a 10 ml bd¿ pre-filled normal saline syringe, in 10 ml syringe.The patient declined to provided further information.*out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 ML BD¿ PRE-FILLED NORMAL SALINE SYRINGE, IN 10 ML SYRINGE
Type of Device
PREFILLED SALINE FLUSH SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
9630 south 54th street
franklin WI 53132
MDR Report Key7571452
MDR Text Key110166721
Report Number2134319-2018-00033
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier30382903065005
UDI-Public30382903065005
Combination Product (y/n)N
PMA/PMN Number
K011982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Expiration Date06/27/2020
Device Catalogue Number306500
Device Lot Number717911C
Was Device Available for Evaluation? No
Date Manufacturer Received05/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-