Catalog Number 328468
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/11/2018
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported by a consumer that the needle hub assembly comes away from syringe after injection when using a bd insulin syringe with the bd ultra-fine¿ needle.There was no report of exposure, serious injury, or medical intervention.
 
Manufacturer Narrative
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch# 7177894.All inspections and challenges were performed per the applicable operations qc specifications.There was one notification noted for raised hubs.There were five notifications noted that did not pertain to the complaint.Based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Type of Device
INSULIN SYRINGE WITH NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key7571578
MDR Text Key110473321
Report Number1920898-2018-00370
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00382908468039
UDI-Public00382908468039
Combination Product (y/n)N
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328468
Device Lot Number7177894
Initial Date Manufacturer Received 05/11/2018
Initial Date FDA Received06/05/2018
Supplement Dates Manufacturer Received05/11/2018
Supplement Dates FDA Received06/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-