AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012453-20
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue.There is no indication of a product quality issue with respects to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the mildly calcified and tortuous right coronary artery.The 4.0 x 20 mm nc trek dilatation catheter was advanced to the target lesion for post-dilatation of an unspecified stent.The dilatation catheter balloon was inflated and no issues were noted; however, during deflation, the balloon would not fully deflate.The balloon was inflated and deflated a few times and the balloon finally deflated enough to be removed from the patient anatomy.There was no difficulty noted during removal from the patient anatomy.There was no reported adverse patient effect or a clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7571769
Report Number2024168-2018-04175
Device Sequence Number1813371
Product Code LOX
UDI-Device Identifier08717648152085
UDI-Public08717648152085
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Catalogue Number1012453-20
Device Lot Number80213G1
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2018
Initial Report FDA Received Date06/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-