| Model Number |
980 |
| Medical Device Problem Code |
Device Inoperable (1663)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Covidien /medtronic has not received the device component from the customer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, while in use on a patient, a 980 ventilator display turned off, alarmed constantly and stopped cycling while in pressure support mode.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the field service engineer (fse) evaluated the device and replaced the graphical user interface (gui), breath delivery (bd) central processing unit (cpu) and oxygen sensor.The ventilator passed all testing per manufacturing specification and was placed back into clinical use.If information is provided in the future, a supplemental report will be issued.
|
| |
|