• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES EB010, VOYANT 5MM FUSION 37CM, 6/BX; GEI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES EB010, VOYANT 5MM FUSION 37CM, 6/BX; GEI Back to Search Results
Model Number EB010
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: the event unit was returned to applied medical for evaluation.Engineering observed that the trigger lock, a metal component located at the bottom of the trigger, was deformed.Based on the condition of the returned unit, it is likely that the reported event was caused by deformation of the trigger lock, which is used to engage and disengage the trigger from the handle.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.Applied medical continuously seeks to improve the form, function and ease of use of its products.As part of this process, applied medical has recently implemented component enhancements intended to further minimize the potential for this type of event to occur.This report reflects the combined initial and final report.
 
Event or Problem Description
Procedure performed: lap.Hysterectomy device worked well ( good sealing) , surgeon was satisfied, except for maniability and ergonomy.She was a bit stressed and probably pressed the handle too strongly since the beginning.At a time, the jaws kept closed on tissue.5-6 tries before it opens again.No injury to the patient.Another device was used to finish the procedure patient status: did a patient injury or illness occur associated with the complaint event? no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB010, VOYANT 5MM FUSION 37CM, 6/BX
Common Device Name
GEI
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key7571911
Report Number2027111-2018-00199
Device Sequence Number11144052
Product Code GEI
UDI-Device Identifier00607915117856
UDI-Public(01)00607915117856(17)190125(30)01(10)1301683
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K141288
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/25/2019
Device Model NumberEB010
Device Catalogue Number101462801
Device Lot Number1301683
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2018
Initial Report FDA Received Date06/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-