| Model Number |
R502-26 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided the age of the device was calculated as the time elapsed between device sterilization and the date of the event.(b)(4).Sorin group (b)(4) manufactures the stockert-shiley aortic root cannulae.The incident occurred in (b)(6).Per exemption number e2016005.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The involved device has been requested for return to sorin group (b)(4) for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not yet returned.
|
| |
|
Event or Problem Description
|
|
Sorin group (b)(4) has received a report that at the beginning of a procedure, a plastic part of the aortic needle broke into two pieces.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Sorin group italia manufactures the stockert-shiley aortic root cannulae.The incident occurred in sassari, italy.Per exemption number e2016005, sorin group italia s.R.L.Is submitting the report for both sorin group italia s.R.L (manufacturer) and livanova usa., inc.(importer).The involved cannula was requested for return to sorin group italia, however the device has not been provided for investigation.During follow-up communication with the customer, photographic were provided which enabled sorin group italia to verify that the larger ring was broken along its longitudinal axis.No other complaint has been received for the reported lot of cannula.Review of the sorin group italia complaint database showed that the reported failure has a very low frequency of occurrence.Based on this, and because the product was not returned for investigation, no corrective actions were identified.Sorin group italia will continue to monitor the market for similar cases.
|
| |
|
Search Alerts/Recalls
|