• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US 9MM X 0DEG X 11MM X 28MM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US 9MM X 0DEG X 11MM X 28MM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 48953090
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Injury (2348); Foreign Body In Patient (2687)
Date of Event 05/14/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that 2 screws pulled out of vertebral body and interbody cage migrated from original position.Patient required revision surgery.
 
Event or Problem Description
It was reported that 2 screws pulled out of vertebral body and interbody cage migrated from original position.Patient required revision surgery.
 
Additional Manufacturer Narrative
Method: device history review, complaint history review, risk assessment; result: the customer reported event was confirmed via correspondence.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: the root cause of the reported event is due to the supplemental fixation system (screws) pulling out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9MM X 0DEG X 11MM X 28MM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7573021
Report Number3004024955-2018-00033
Device Sequence Number18886421
Product Code MAX
UDI-Device Identifier07613327117851
UDI-Public(01)07613327117851
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160955
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48953090
Device Catalogue Number48953090
Device Lot NumberD58N
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/14/2018
Supplement Date Received by Manufacturer08/30/2018
Initial Report FDA Received Date06/06/2018
Supplement Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-