| Model Number |
488516535 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Injury (2348); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/14/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that 2 screws pulled out of vertebral body and interbody cage migrated from original position.Patient required revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: risk assessment.The event was confirmed based communication with sales representative.The screw was removed but not replaced and are unavailable for evaluation.X-rays were not provided.The lot number is unknown, thus, a manufacturing records review could not be completed.As the device and x rays are not available, a definite root cause cannot be identified.It is possible the screw was placed too proud in the bone, which could reduce bone purchase and cause pullout.Patient bone quality could also be a factor as poor bone quality can affect screw purchase.Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that 2 screws pulled out of vertebral body and interbody cage migrated from original position.Patient required revision surgery.
|
| |
|
Search Alerts/Recalls
|