MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Kinked (1339); Unstable (1667)
|
Patient Problem
Therapeutic Response, Decreased (2271)
|
Event Date 05/25/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Device code (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2017, product type: catheter.Product id: 8780, serial/lot #: (b)(4), ubd: 10-oct-2019, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a health care provider (hcp) via a device manufacturer representative regarding a patient receiving morphine (10 mg/ml at 1 mg/day) via an implantable infusion pump.The indication for use was noted to be spinal pain.It was reported that a refill was attempted but the pump was flipped.This pump flip caused a kinked catheter which led to a loss of therapy.The event date was noted to be (b)(6) 2018.The kinked portion of the catheter was removed and replaced, and normal cerebrospinal fluid flow was obtained.The issue was considered resolved.The patient's status at the time of the report was alive - no injury.No further complications were reported.
|
|