• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Kinked (1339); Unstable (1667)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 05/25/2018
Event Type  Injury  
Manufacturer Narrative
Device code (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2017, product type: catheter.Product id: 8780, serial/lot #: (b)(4), ubd: 10-oct-2019, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care provider (hcp) via a device manufacturer representative regarding a patient receiving morphine (10 mg/ml at 1 mg/day) via an implantable infusion pump.The indication for use was noted to be spinal pain.It was reported that a refill was attempted but the pump was flipped.This pump flip caused a kinked catheter which led to a loss of therapy.The event date was noted to be (b)(6) 2018.The kinked portion of the catheter was removed and replaced, and normal cerebrospinal fluid flow was obtained.The issue was considered resolved.The patient's status at the time of the report was alive - no injury.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7573374
MDR Text Key110216658
Report Number3004209178-2018-12707
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530119
UDI-Public00643169530119
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2019
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/04/2018
Date Device Manufactured11/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient Weight75
-
-