| Catalog Number |
9734716K |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No parts have been received by the manufacturer for evaluation as the site declined to replace the instrument.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a mast/perc spinal fusion.It was reported that intra-operatively, while setting up equipment.The short hex screw spinous process clamp would close when the clamp was held in the air, but not when positioned around a spinous process.Surgery was completed using the navigation system and a different set of spinous process clamps.Surgical time was extended by less than one hour and there was no reported patient impact.
|
| |
|
Search Alerts/Recalls
|