| Model Number |
8900-2303-01 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the associated device displayed a "check pads" message using these electrode pads.Complainant indicated that the clinician obtained several sets of electrode pads (one same lot 1518 and one different lot 1218 - manufacture date of 3/19/2018 expiration date of 12/19/2019) to continue treating the patient and the "check pads" message appeared.The complainant indicated that this caused a delay in the treatment of the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual electrodes were not returned for evaluation.Instead, an un-opened set of electrode pads were returned.The reported problem could not be verified.The un-opened electrode pads returned passed visual inspection with no discrepancies noted.Two lots 1518 and 1218 were pulled for retained sample testing.None of the units resulted in a check pads message when subjected to simulated testing.Analysis of reports of this type has not identified an increase in trend.
|
| |
|