Model Number 304-20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Muscle Spasm(s) (1966); Neck Pain (2433)
Event Date 11/01/2017
Event Type  Injury  
Event Description
It was reported that the patient is experiencing muscle spasms with vns stimulation.When the device stimulates, patient's muscle by the neck twitches.All of the diagnostics were within normal limits.Patient underwent parathyroid surgery and since then, the patient has had bulging of the muscle over the left side of his neck with vns stimulation.Since reducing the output current , the bulging was reported to be tolerable.Patient underwent lead revision surgery due to painful stimulation in the muscle.The explanted lead was received.Analysis is underway but has not been completed to date.
 
Event Description
A portion of the lead assembly (body) including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The set screw marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.Based on the findings in the product analysis lab, there is no evidence to suggest an anomaly with the returned portions of the device which may have contributed to the stated complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7574456
MDR Text Key110256470
Report Number1644487-2018-00970
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750139
UDI-Public05425025750139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/30/2014
Device Model Number304-20
Device Lot Number2639
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received06/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age47 YR
-
-