Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and evaluated.The device was observed under a microscope and confirmed there are nicks, slight discoloration and missing pieces of metal found at the opening near the distal tip of the shaft.The device appears in used condition.The friction marks indicate excessive force caused by the lack of irrigation fluid, therefore this failure can be attributed to inadequate amount of fluid used.A device history record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.Further, a review into the depuy synthes mitek complaints system revealed no other complaints of any kind for this lot of devices with the product code that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
|