| Model Number |
H7493952820400 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 4.00x20mm promus premier¿ stent was advanced for treatment and was successfully deployed.However, when the stent delivery system was removed post-stent deployment, it was noted that the shaft was broken.No patient complications were reported and the patient's status was fine.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 4.00x20mm promus premier¿ stent was advanced for treatment and was successfully deployed.However, when the stent delivery system was removed post-stent deployment, it was noted that the shaft was broken.No patient complications were reported and the patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr.: promus premier us mr 2.25 x 32mm stent delivery system was returned for analysis.The stent was not returned as it was deployed during the procedure.The balloon was examined; the balloon folds were opened.The balloon appeared to have been inflated and deflated.The distal end of the balloon appeared to be bunched.A leak test was carried out: the balloon was inflated to rated burst pressure, there were no issues with the inflation.There were no leaks noted in the device.A vacuum was pulled and the balloon deflated fully in 3s, which is within the specified time.A visual and tactile examination of the device found multiple hypotube kinks.An examination of the shaft polymer extrusion identified a bunch in the inner shaft polymer extrusion at 2mm distal of the bi-component bond.The tip was visually examined and no issues were noted.No other issues were identified during the product analysis.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|